A prototype instrument for single pinhole small animal adaptive SPECT imaging

Melanie Freed, Matthew A Kupinski, Lars R Furenlid, Donald W. Wilson, Harrison H Barrett

Research output: Contribution to journalArticle

39 Citations (Scopus)

Abstract

The authors have designed and constructed a small-animal adaptive SPECT imaging system as a prototype for quantifying the potential benefit of adaptive SPECT imaging over the traditional fixed geometry approach. The optical design of the system is based on filling the detector with the region of interest for each viewing angle, maximizing the sensitivity, and optimizing the resolution in the projection images. Additional feedback rules for determining the optimal geometry of the system can be easily added to the existing control software. Preliminary data have been taken of a phantom with a small, hot, offset lesion in a flat background in both adaptive and fixed geometry modes. Comparison of the predicted system behavior with the actual system behavior is presented, along with recommendations for system improvements.

Original languageEnglish (US)
Pages (from-to)1912-1925
Number of pages14
JournalMedical Physics
Volume35
Issue number5
DOIs
StatePublished - 2008

Fingerprint

Single-Photon Emission-Computed Tomography
Optical Devices
Software

Keywords

  • Adaptive imaging
  • Instrumentation
  • Optimization
  • Small-animal SPECT

ASJC Scopus subject areas

  • Biophysics

Cite this

A prototype instrument for single pinhole small animal adaptive SPECT imaging. / Freed, Melanie; Kupinski, Matthew A; Furenlid, Lars R; Wilson, Donald W.; Barrett, Harrison H.

In: Medical Physics, Vol. 35, No. 5, 2008, p. 1912-1925.

Research output: Contribution to journalArticle

@article{a4e80d238c914844838d96666cb03fe0,
title = "A prototype instrument for single pinhole small animal adaptive SPECT imaging",
abstract = "The authors have designed and constructed a small-animal adaptive SPECT imaging system as a prototype for quantifying the potential benefit of adaptive SPECT imaging over the traditional fixed geometry approach. The optical design of the system is based on filling the detector with the region of interest for each viewing angle, maximizing the sensitivity, and optimizing the resolution in the projection images. Additional feedback rules for determining the optimal geometry of the system can be easily added to the existing control software. Preliminary data have been taken of a phantom with a small, hot, offset lesion in a flat background in both adaptive and fixed geometry modes. Comparison of the predicted system behavior with the actual system behavior is presented, along with recommendations for system improvements.",
keywords = "Adaptive imaging, Instrumentation, Optimization, Small-animal SPECT",
author = "Melanie Freed and Kupinski, {Matthew A} and Furenlid, {Lars R} and Wilson, {Donald W.} and Barrett, {Harrison H}",
year = "2008",
doi = "10.1118/1.2896072",
language = "English (US)",
volume = "35",
pages = "1912--1925",
journal = "Medical Physics",
issn = "0094-2405",
publisher = "AAPM - American Association of Physicists in Medicine",
number = "5",

}

TY - JOUR

T1 - A prototype instrument for single pinhole small animal adaptive SPECT imaging

AU - Freed, Melanie

AU - Kupinski, Matthew A

AU - Furenlid, Lars R

AU - Wilson, Donald W.

AU - Barrett, Harrison H

PY - 2008

Y1 - 2008

N2 - The authors have designed and constructed a small-animal adaptive SPECT imaging system as a prototype for quantifying the potential benefit of adaptive SPECT imaging over the traditional fixed geometry approach. The optical design of the system is based on filling the detector with the region of interest for each viewing angle, maximizing the sensitivity, and optimizing the resolution in the projection images. Additional feedback rules for determining the optimal geometry of the system can be easily added to the existing control software. Preliminary data have been taken of a phantom with a small, hot, offset lesion in a flat background in both adaptive and fixed geometry modes. Comparison of the predicted system behavior with the actual system behavior is presented, along with recommendations for system improvements.

AB - The authors have designed and constructed a small-animal adaptive SPECT imaging system as a prototype for quantifying the potential benefit of adaptive SPECT imaging over the traditional fixed geometry approach. The optical design of the system is based on filling the detector with the region of interest for each viewing angle, maximizing the sensitivity, and optimizing the resolution in the projection images. Additional feedback rules for determining the optimal geometry of the system can be easily added to the existing control software. Preliminary data have been taken of a phantom with a small, hot, offset lesion in a flat background in both adaptive and fixed geometry modes. Comparison of the predicted system behavior with the actual system behavior is presented, along with recommendations for system improvements.

KW - Adaptive imaging

KW - Instrumentation

KW - Optimization

KW - Small-animal SPECT

UR - http://www.scopus.com/inward/record.url?scp=42949165863&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=42949165863&partnerID=8YFLogxK

U2 - 10.1118/1.2896072

DO - 10.1118/1.2896072

M3 - Article

C2 - 18561667

AN - SCOPUS:42949165863

VL - 35

SP - 1912

EP - 1925

JO - Medical Physics

JF - Medical Physics

SN - 0094-2405

IS - 5

ER -