A real-time reconstruction system for magnetic resonance imaging

Arthur F Gmitro, Ali R. Ehsani, Troy A. Berchem, Rodney J. Snell

Research output: Contribution to journalArticle

21 Citations (Scopus)

Abstract

A digital-electronic reconstruction system for MRI has been designed and demonstrated. The system is capable of reconstructing a 128 x 128 pixel image from complex-valued data in approximately 8 ms (122 frames per second) or a 256 x 256 pixel image in 32 ms (30 frames per second) using the standard 2D FFT reconstruction algorithm. Real-time MR imaging can be obtained when this reconstruction system is coupled with fast continuous echo-planar type data acquisition. This provides the unique potential for real-time monitoring of interventional procedures or for rapid patient positioning. The real-time reconstruction system presented here consists of four main subsystems: an analog to digital converter, an interface memory, the Fourier processor, and the display processor. The basic design of this reconstruction system is presented along with results, demonstrating the capability of the system.

Original languageEnglish (US)
Pages (from-to)734-740
Number of pages7
JournalMagnetic Resonance in Medicine
Volume35
Issue number5
DOIs
StatePublished - May 1996

Fingerprint

Computer Systems
Magnetic Resonance Imaging
Patient Positioning

Keywords

  • hardware
  • image reconstruction
  • MRI
  • real-time

ASJC Scopus subject areas

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology

Cite this

A real-time reconstruction system for magnetic resonance imaging. / Gmitro, Arthur F; Ehsani, Ali R.; Berchem, Troy A.; Snell, Rodney J.

In: Magnetic Resonance in Medicine, Vol. 35, No. 5, 05.1996, p. 734-740.

Research output: Contribution to journalArticle

Gmitro, Arthur F ; Ehsani, Ali R. ; Berchem, Troy A. ; Snell, Rodney J. / A real-time reconstruction system for magnetic resonance imaging. In: Magnetic Resonance in Medicine. 1996 ; Vol. 35, No. 5. pp. 734-740.
@article{fdf65ebdd3ac4ec59674270b93234c7d,
title = "A real-time reconstruction system for magnetic resonance imaging",
abstract = "A digital-electronic reconstruction system for MRI has been designed and demonstrated. The system is capable of reconstructing a 128 x 128 pixel image from complex-valued data in approximately 8 ms (122 frames per second) or a 256 x 256 pixel image in 32 ms (30 frames per second) using the standard 2D FFT reconstruction algorithm. Real-time MR imaging can be obtained when this reconstruction system is coupled with fast continuous echo-planar type data acquisition. This provides the unique potential for real-time monitoring of interventional procedures or for rapid patient positioning. The real-time reconstruction system presented here consists of four main subsystems: an analog to digital converter, an interface memory, the Fourier processor, and the display processor. The basic design of this reconstruction system is presented along with results, demonstrating the capability of the system.",
keywords = "hardware, image reconstruction, MRI, real-time",
author = "Gmitro, {Arthur F} and Ehsani, {Ali R.} and Berchem, {Troy A.} and Snell, {Rodney J.}",
year = "1996",
month = "5",
doi = "10.1002/mrm.1910350515",
language = "English (US)",
volume = "35",
pages = "734--740",
journal = "Magnetic Resonance in Medicine",
issn = "0740-3194",
publisher = "John Wiley and Sons Inc.",
number = "5",

}

TY - JOUR

T1 - A real-time reconstruction system for magnetic resonance imaging

AU - Gmitro, Arthur F

AU - Ehsani, Ali R.

AU - Berchem, Troy A.

AU - Snell, Rodney J.

PY - 1996/5

Y1 - 1996/5

N2 - A digital-electronic reconstruction system for MRI has been designed and demonstrated. The system is capable of reconstructing a 128 x 128 pixel image from complex-valued data in approximately 8 ms (122 frames per second) or a 256 x 256 pixel image in 32 ms (30 frames per second) using the standard 2D FFT reconstruction algorithm. Real-time MR imaging can be obtained when this reconstruction system is coupled with fast continuous echo-planar type data acquisition. This provides the unique potential for real-time monitoring of interventional procedures or for rapid patient positioning. The real-time reconstruction system presented here consists of four main subsystems: an analog to digital converter, an interface memory, the Fourier processor, and the display processor. The basic design of this reconstruction system is presented along with results, demonstrating the capability of the system.

AB - A digital-electronic reconstruction system for MRI has been designed and demonstrated. The system is capable of reconstructing a 128 x 128 pixel image from complex-valued data in approximately 8 ms (122 frames per second) or a 256 x 256 pixel image in 32 ms (30 frames per second) using the standard 2D FFT reconstruction algorithm. Real-time MR imaging can be obtained when this reconstruction system is coupled with fast continuous echo-planar type data acquisition. This provides the unique potential for real-time monitoring of interventional procedures or for rapid patient positioning. The real-time reconstruction system presented here consists of four main subsystems: an analog to digital converter, an interface memory, the Fourier processor, and the display processor. The basic design of this reconstruction system is presented along with results, demonstrating the capability of the system.

KW - hardware

KW - image reconstruction

KW - MRI

KW - real-time

UR - http://www.scopus.com/inward/record.url?scp=0029916585&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029916585&partnerID=8YFLogxK

U2 - 10.1002/mrm.1910350515

DO - 10.1002/mrm.1910350515

M3 - Article

C2 - 8722825

AN - SCOPUS:0029916585

VL - 35

SP - 734

EP - 740

JO - Magnetic Resonance in Medicine

JF - Magnetic Resonance in Medicine

SN - 0740-3194