Accuracy of the PAR Comeal Topography System with spatial misalignment

Michael W Belin, Peter Zloty

Research output: Contribution to journalArticle

29 Citations (Scopus)

Abstract

The PAR Corneal Topography System is a computerized corneal imaging system which uses close-range raster photogrammetry to measure and produce a topographic map of the corneal surface. Raster photogrammetry is a standard method of extracting object information by projecting a known pattern onto an object and recording the distortion when viewed from an oblique angle. Unlike placido disc based videokeratoscopes, the PAR system requires neither a smooth reflective surface nor precise spatial alignment for accurate imaging. We studied both the accuracy of the system with purposeful misalignment (defocusing) of the test object and determined the ability to image freshly deepithelialized, keratectomized, and photoablated corneas. The PAR system was both accurate and reproducible in imaging calibrated spheres within a defined zone in space. Whole cadaver eyes were imaged both before and immediately after removal of the epithelium, lamellar keratectomy, and laser photoablation. The system demonstrated the ability to image irregular, deepithelialized, and keratectomized corneas. The ability to maintain accuracy without precise alignment and the facility to image freshly deepithelialized and keratectomized corneas may make the system suitable for intraoperative refractive monitoring.

Original languageEnglish (US)
Pages (from-to)64-68
Number of pages5
JournalEye and Contact Lens
Volume19
Issue number1
StatePublished - 1993
Externally publishedYes

Fingerprint

Photogrammetry
Cornea
Laser Corneal Surgery
Corneal Topography
Intraoperative Monitoring
Cadaver
Epithelium

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Accuracy of the PAR Comeal Topography System with spatial misalignment. / Belin, Michael W; Zloty, Peter.

In: Eye and Contact Lens, Vol. 19, No. 1, 1993, p. 64-68.

Research output: Contribution to journalArticle

@article{b1729dc6e2414be0b699dbbcd1ee7632,
title = "Accuracy of the PAR Comeal Topography System with spatial misalignment",
abstract = "The PAR Corneal Topography System is a computerized corneal imaging system which uses close-range raster photogrammetry to measure and produce a topographic map of the corneal surface. Raster photogrammetry is a standard method of extracting object information by projecting a known pattern onto an object and recording the distortion when viewed from an oblique angle. Unlike placido disc based videokeratoscopes, the PAR system requires neither a smooth reflective surface nor precise spatial alignment for accurate imaging. We studied both the accuracy of the system with purposeful misalignment (defocusing) of the test object and determined the ability to image freshly deepithelialized, keratectomized, and photoablated corneas. The PAR system was both accurate and reproducible in imaging calibrated spheres within a defined zone in space. Whole cadaver eyes were imaged both before and immediately after removal of the epithelium, lamellar keratectomy, and laser photoablation. The system demonstrated the ability to image irregular, deepithelialized, and keratectomized corneas. The ability to maintain accuracy without precise alignment and the facility to image freshly deepithelialized and keratectomized corneas may make the system suitable for intraoperative refractive monitoring.",
author = "Belin, {Michael W} and Peter Zloty",
year = "1993",
language = "English (US)",
volume = "19",
pages = "64--68",
journal = "Eye and Contact Lense",
issn = "1542-2321",
publisher = "Lippincott Williams and Wilkins",
number = "1",

}

TY - JOUR

T1 - Accuracy of the PAR Comeal Topography System with spatial misalignment

AU - Belin, Michael W

AU - Zloty, Peter

PY - 1993

Y1 - 1993

N2 - The PAR Corneal Topography System is a computerized corneal imaging system which uses close-range raster photogrammetry to measure and produce a topographic map of the corneal surface. Raster photogrammetry is a standard method of extracting object information by projecting a known pattern onto an object and recording the distortion when viewed from an oblique angle. Unlike placido disc based videokeratoscopes, the PAR system requires neither a smooth reflective surface nor precise spatial alignment for accurate imaging. We studied both the accuracy of the system with purposeful misalignment (defocusing) of the test object and determined the ability to image freshly deepithelialized, keratectomized, and photoablated corneas. The PAR system was both accurate and reproducible in imaging calibrated spheres within a defined zone in space. Whole cadaver eyes were imaged both before and immediately after removal of the epithelium, lamellar keratectomy, and laser photoablation. The system demonstrated the ability to image irregular, deepithelialized, and keratectomized corneas. The ability to maintain accuracy without precise alignment and the facility to image freshly deepithelialized and keratectomized corneas may make the system suitable for intraoperative refractive monitoring.

AB - The PAR Corneal Topography System is a computerized corneal imaging system which uses close-range raster photogrammetry to measure and produce a topographic map of the corneal surface. Raster photogrammetry is a standard method of extracting object information by projecting a known pattern onto an object and recording the distortion when viewed from an oblique angle. Unlike placido disc based videokeratoscopes, the PAR system requires neither a smooth reflective surface nor precise spatial alignment for accurate imaging. We studied both the accuracy of the system with purposeful misalignment (defocusing) of the test object and determined the ability to image freshly deepithelialized, keratectomized, and photoablated corneas. The PAR system was both accurate and reproducible in imaging calibrated spheres within a defined zone in space. Whole cadaver eyes were imaged both before and immediately after removal of the epithelium, lamellar keratectomy, and laser photoablation. The system demonstrated the ability to image irregular, deepithelialized, and keratectomized corneas. The ability to maintain accuracy without precise alignment and the facility to image freshly deepithelialized and keratectomized corneas may make the system suitable for intraoperative refractive monitoring.

UR - http://www.scopus.com/inward/record.url?scp=0027536610&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0027536610&partnerID=8YFLogxK

M3 - Article

C2 - 8453756

AN - SCOPUS:0027536610

VL - 19

SP - 64

EP - 68

JO - Eye and Contact Lense

JF - Eye and Contact Lense

SN - 1542-2321

IS - 1

ER -