Comparison of myocardial damage from defibrillator discharges at various dosages

G. A. Ewy, Douglas L Taren, J. Bangert, S. McClung, D. A. Hellman

Research output: Contribution to journalArticle

18 Citations (Scopus)

Abstract

To study the difference in myocardial damage produced by the same defibrillation energy delivered at frequent, low-energy discharges vs infrequent, high-energy discharges, experiments were performed on 48 dogs. In Part I, a total of 3000 Wsec energy was delivered; in Part II, a total of 600 Wsec was delivered; and in Parts III and IV, a total of 1800 Wsec was delivered. In half the dogs, the energy was delivered in three shocks within 135 sec. In the other half, the shocks were delivered by 10 smaller shocks at 15-sec intervals. The degree of myocardial damage assessed by St-segment elevation on precordial mapping or necrotic index from gross and microscopic examination 4 days after the shocks showed no significant difference. However, the three high-energy shocks (600 to 1000 Wsec) resulted in ventricular fibrillation or complete heart block in 33% (6 of 18) of the dogs. These arrhythmias were not observed in the animals receiving low-energy (60-, 200-, or 300-Wsec) discharges.

Original languageEnglish (US)
Pages (from-to)9-12
Number of pages4
JournalMedical Instrumentation
Volume14
Issue number1
StatePublished - Jan 1980

Fingerprint

Defibrillators
Shock
Microscopic examination
Animals
Dogs
Experiments
Heart Block
Ventricular Fibrillation
Cardiac Arrhythmias

ASJC Scopus subject areas

  • Medicine (miscellaneous)
  • Bioengineering

Cite this

Ewy, G. A., Taren, D. L., Bangert, J., McClung, S., & Hellman, D. A. (1980). Comparison of myocardial damage from defibrillator discharges at various dosages. Medical Instrumentation, 14(1), 9-12.

Comparison of myocardial damage from defibrillator discharges at various dosages. / Ewy, G. A.; Taren, Douglas L; Bangert, J.; McClung, S.; Hellman, D. A.

In: Medical Instrumentation, Vol. 14, No. 1, 01.1980, p. 9-12.

Research output: Contribution to journalArticle

Ewy, GA, Taren, DL, Bangert, J, McClung, S & Hellman, DA 1980, 'Comparison of myocardial damage from defibrillator discharges at various dosages', Medical Instrumentation, vol. 14, no. 1, pp. 9-12.
Ewy, G. A. ; Taren, Douglas L ; Bangert, J. ; McClung, S. ; Hellman, D. A. / Comparison of myocardial damage from defibrillator discharges at various dosages. In: Medical Instrumentation. 1980 ; Vol. 14, No. 1. pp. 9-12.
@article{0c009f6d36134fefa7463474c9e6775b,
title = "Comparison of myocardial damage from defibrillator discharges at various dosages",
abstract = "To study the difference in myocardial damage produced by the same defibrillation energy delivered at frequent, low-energy discharges vs infrequent, high-energy discharges, experiments were performed on 48 dogs. In Part I, a total of 3000 Wsec energy was delivered; in Part II, a total of 600 Wsec was delivered; and in Parts III and IV, a total of 1800 Wsec was delivered. In half the dogs, the energy was delivered in three shocks within 135 sec. In the other half, the shocks were delivered by 10 smaller shocks at 15-sec intervals. The degree of myocardial damage assessed by St-segment elevation on precordial mapping or necrotic index from gross and microscopic examination 4 days after the shocks showed no significant difference. However, the three high-energy shocks (600 to 1000 Wsec) resulted in ventricular fibrillation or complete heart block in 33{\%} (6 of 18) of the dogs. These arrhythmias were not observed in the animals receiving low-energy (60-, 200-, or 300-Wsec) discharges.",
author = "Ewy, {G. A.} and Taren, {Douglas L} and J. Bangert and S. McClung and Hellman, {D. A.}",
year = "1980",
month = "1",
language = "English (US)",
volume = "14",
pages = "9--12",
journal = "Biomedical Instrumentation and Technology",
issn = "0899-8205",
publisher = "Association for the Advancement of Medical Instrumentation",
number = "1",

}

TY - JOUR

T1 - Comparison of myocardial damage from defibrillator discharges at various dosages

AU - Ewy, G. A.

AU - Taren, Douglas L

AU - Bangert, J.

AU - McClung, S.

AU - Hellman, D. A.

PY - 1980/1

Y1 - 1980/1

N2 - To study the difference in myocardial damage produced by the same defibrillation energy delivered at frequent, low-energy discharges vs infrequent, high-energy discharges, experiments were performed on 48 dogs. In Part I, a total of 3000 Wsec energy was delivered; in Part II, a total of 600 Wsec was delivered; and in Parts III and IV, a total of 1800 Wsec was delivered. In half the dogs, the energy was delivered in three shocks within 135 sec. In the other half, the shocks were delivered by 10 smaller shocks at 15-sec intervals. The degree of myocardial damage assessed by St-segment elevation on precordial mapping or necrotic index from gross and microscopic examination 4 days after the shocks showed no significant difference. However, the three high-energy shocks (600 to 1000 Wsec) resulted in ventricular fibrillation or complete heart block in 33% (6 of 18) of the dogs. These arrhythmias were not observed in the animals receiving low-energy (60-, 200-, or 300-Wsec) discharges.

AB - To study the difference in myocardial damage produced by the same defibrillation energy delivered at frequent, low-energy discharges vs infrequent, high-energy discharges, experiments were performed on 48 dogs. In Part I, a total of 3000 Wsec energy was delivered; in Part II, a total of 600 Wsec was delivered; and in Parts III and IV, a total of 1800 Wsec was delivered. In half the dogs, the energy was delivered in three shocks within 135 sec. In the other half, the shocks were delivered by 10 smaller shocks at 15-sec intervals. The degree of myocardial damage assessed by St-segment elevation on precordial mapping or necrotic index from gross and microscopic examination 4 days after the shocks showed no significant difference. However, the three high-energy shocks (600 to 1000 Wsec) resulted in ventricular fibrillation or complete heart block in 33% (6 of 18) of the dogs. These arrhythmias were not observed in the animals receiving low-energy (60-, 200-, or 300-Wsec) discharges.

UR - http://www.scopus.com/inward/record.url?scp=0018862879&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0018862879&partnerID=8YFLogxK

M3 - Article

C2 - 7354738

AN - SCOPUS:0018862879

VL - 14

SP - 9

EP - 12

JO - Biomedical Instrumentation and Technology

JF - Biomedical Instrumentation and Technology

SN - 0899-8205

IS - 1

ER -