Demountable 4He film stopper for dilution refrigerators

B. M. Abraham, Charles M. Falco

Research output: Contribution to journalArticle

6 Scopus citations

Abstract

We describe an easily assembled and demountable film stopper for use in either a pumped 4He pot or a 3He-4He dilution refrigerator. The orifice plate for the film stopper is fabricated from a Pyrex glass plate and is sealed to the metal body of the still with an indium O-ring. The advantage of glass over metal for the orifice plate is achieved without complicated construction.

Original languageEnglish (US)
Pages (from-to)253-254
Number of pages2
JournalReview of Scientific Instruments
Volume47
Issue number2
DOIs
StatePublished - Dec 1 1976