In this report we describe the outcome of a 6-week study of the valve- stent device in a goat model. Six female goats underwent endoscopic subcutaneous harvest of a valve-containing segment of external jugular vein. This segment was sutured inside a 10-mm Wallstent (Schneider Inc., USA), and the device was compressed and loaded into a 12 Fr. introducer sheath and deployed through a venotomy into the contralateral external jugular vein. The animals were anticoagulated for 1 week and euthanized at 6 weeks. Flow studies, manual strip test, and angioscopic evaluation were performed prior to sacrifice. All six valve-stents were patent at harvest by Doppler flow studies, and five of six were competent as demonstrated by the manual strip test. At angioscopic evaluation, the valves were open. Gross and histologic examination revealed intact valve leaflets in five specimens. This technique is a minimally invasive approach to valve harvest and transplantation with a valve-stent device. We have demonstrated that valve-stent can remain patent and competent for up to 6 weeks. This technique may have potential application in venous valvular incompetence.

Original languageEnglish (US)
Pages (from-to)62-67
Number of pages6
JournalAnnals of Vascular Surgery
Volume11
Issue number1
DOIs
StatePublished - Jan 1997
Externally publishedYes

Fingerprint

Venous Valves
Stents
Transplantation
Equipment and Supplies
Jugular Veins
Goats

ASJC Scopus subject areas

  • Cardiology and Cardiovascular Medicine

Cite this

Endoscopic venous valve transplantation with a valve-stent device. / Ofenloch, John C.; Chen, Changyi; Hughes, John D; Lumsden, Alan B.

In: Annals of Vascular Surgery, Vol. 11, No. 1, 01.1997, p. 62-67.

Research output: Contribution to journalArticle

Ofenloch, John C. ; Chen, Changyi ; Hughes, John D ; Lumsden, Alan B. / Endoscopic venous valve transplantation with a valve-stent device. In: Annals of Vascular Surgery. 1997 ; Vol. 11, No. 1. pp. 62-67.
@article{44861727ef224b1c9662b6a138036f78,
title = "Endoscopic venous valve transplantation with a valve-stent device",
abstract = "In this report we describe the outcome of a 6-week study of the valve- stent device in a goat model. Six female goats underwent endoscopic subcutaneous harvest of a valve-containing segment of external jugular vein. This segment was sutured inside a 10-mm Wallstent (Schneider Inc., USA), and the device was compressed and loaded into a 12 Fr. introducer sheath and deployed through a venotomy into the contralateral external jugular vein. The animals were anticoagulated for 1 week and euthanized at 6 weeks. Flow studies, manual strip test, and angioscopic evaluation were performed prior to sacrifice. All six valve-stents were patent at harvest by Doppler flow studies, and five of six were competent as demonstrated by the manual strip test. At angioscopic evaluation, the valves were open. Gross and histologic examination revealed intact valve leaflets in five specimens. This technique is a minimally invasive approach to valve harvest and transplantation with a valve-stent device. We have demonstrated that valve-stent can remain patent and competent for up to 6 weeks. This technique may have potential application in venous valvular incompetence.",
author = "Ofenloch, {John C.} and Changyi Chen and Hughes, {John D} and Lumsden, {Alan B.}",
year = "1997",
month = "1",
doi = "10.1007/s100169900011",
language = "English (US)",
volume = "11",
pages = "62--67",
journal = "Annals of Vascular Surgery",
issn = "0890-5096",
publisher = "Elsevier Inc.",
number = "1",

}

TY - JOUR

T1 - Endoscopic venous valve transplantation with a valve-stent device

AU - Ofenloch, John C.

AU - Chen, Changyi

AU - Hughes, John D

AU - Lumsden, Alan B.

PY - 1997/1

Y1 - 1997/1

N2 - In this report we describe the outcome of a 6-week study of the valve- stent device in a goat model. Six female goats underwent endoscopic subcutaneous harvest of a valve-containing segment of external jugular vein. This segment was sutured inside a 10-mm Wallstent (Schneider Inc., USA), and the device was compressed and loaded into a 12 Fr. introducer sheath and deployed through a venotomy into the contralateral external jugular vein. The animals were anticoagulated for 1 week and euthanized at 6 weeks. Flow studies, manual strip test, and angioscopic evaluation were performed prior to sacrifice. All six valve-stents were patent at harvest by Doppler flow studies, and five of six were competent as demonstrated by the manual strip test. At angioscopic evaluation, the valves were open. Gross and histologic examination revealed intact valve leaflets in five specimens. This technique is a minimally invasive approach to valve harvest and transplantation with a valve-stent device. We have demonstrated that valve-stent can remain patent and competent for up to 6 weeks. This technique may have potential application in venous valvular incompetence.

AB - In this report we describe the outcome of a 6-week study of the valve- stent device in a goat model. Six female goats underwent endoscopic subcutaneous harvest of a valve-containing segment of external jugular vein. This segment was sutured inside a 10-mm Wallstent (Schneider Inc., USA), and the device was compressed and loaded into a 12 Fr. introducer sheath and deployed through a venotomy into the contralateral external jugular vein. The animals were anticoagulated for 1 week and euthanized at 6 weeks. Flow studies, manual strip test, and angioscopic evaluation were performed prior to sacrifice. All six valve-stents were patent at harvest by Doppler flow studies, and five of six were competent as demonstrated by the manual strip test. At angioscopic evaluation, the valves were open. Gross and histologic examination revealed intact valve leaflets in five specimens. This technique is a minimally invasive approach to valve harvest and transplantation with a valve-stent device. We have demonstrated that valve-stent can remain patent and competent for up to 6 weeks. This technique may have potential application in venous valvular incompetence.

UR - http://www.scopus.com/inward/record.url?scp=0030892069&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0030892069&partnerID=8YFLogxK

U2 - 10.1007/s100169900011

DO - 10.1007/s100169900011

M3 - Article

VL - 11

SP - 62

EP - 67

JO - Annals of Vascular Surgery

JF - Annals of Vascular Surgery

SN - 0890-5096

IS - 1

ER -