Evaluating the trace-moisture measurement capability of coulometric hygrometry

Ce Ma, Farhang Shadman, Jacques Mettes, Lance Silverman

Research output: Contribution to journalArticle

1 Citation (Scopus)

Abstract

Study data show that an advanced coulometric hygrometer design incorporating a counterflow sensor cell, a moisture addition feature, and signal integration extends the range of the measurements technique to single-digit ppd levels. Characterization of the hygrometer in the integration mode in parallel with an atmospheric pressure ionization mass spectrometer (APIMS) indicated that its accuracy is ±4 ppb, its detection limit is 1 ppd at a 95% confidence limit, and its response time to a 5-ppb moisture intrusion is 50% in 15 minutes.

Original languageEnglish (US)
Pages (from-to)43-49
Number of pages7
JournalMICRO
Volume13
Issue number4
StatePublished - Apr 1995

Fingerprint

Hygrometers
hygrometers
moisture
Moisture
confidence limits
counterflow
digits
Mass spectrometers
intrusion
mass spectrometers
Atmospheric pressure
Ionization
atmospheric pressure
ionization
sensors
Sensors
cells

ASJC Scopus subject areas

  • Engineering(all)

Cite this

Evaluating the trace-moisture measurement capability of coulometric hygrometry. / Ma, Ce; Shadman, Farhang; Mettes, Jacques; Silverman, Lance.

In: MICRO, Vol. 13, No. 4, 04.1995, p. 43-49.

Research output: Contribution to journalArticle

Ma, C, Shadman, F, Mettes, J & Silverman, L 1995, 'Evaluating the trace-moisture measurement capability of coulometric hygrometry', MICRO, vol. 13, no. 4, pp. 43-49.
Ma, Ce ; Shadman, Farhang ; Mettes, Jacques ; Silverman, Lance. / Evaluating the trace-moisture measurement capability of coulometric hygrometry. In: MICRO. 1995 ; Vol. 13, No. 4. pp. 43-49.
@article{aa882cf47019436c99f01dcfa441b7f8,
title = "Evaluating the trace-moisture measurement capability of coulometric hygrometry",
abstract = "Study data show that an advanced coulometric hygrometer design incorporating a counterflow sensor cell, a moisture addition feature, and signal integration extends the range of the measurements technique to single-digit ppd levels. Characterization of the hygrometer in the integration mode in parallel with an atmospheric pressure ionization mass spectrometer (APIMS) indicated that its accuracy is ±4 ppb, its detection limit is 1 ppd at a 95{\%} confidence limit, and its response time to a 5-ppb moisture intrusion is 50{\%} in 15 minutes.",
author = "Ce Ma and Farhang Shadman and Jacques Mettes and Lance Silverman",
year = "1995",
month = "4",
language = "English (US)",
volume = "13",
pages = "43--49",
journal = "MICRO",
issn = "1081-0595",
publisher = "Canon Communications LLC",
number = "4",

}

TY - JOUR

T1 - Evaluating the trace-moisture measurement capability of coulometric hygrometry

AU - Ma, Ce

AU - Shadman, Farhang

AU - Mettes, Jacques

AU - Silverman, Lance

PY - 1995/4

Y1 - 1995/4

N2 - Study data show that an advanced coulometric hygrometer design incorporating a counterflow sensor cell, a moisture addition feature, and signal integration extends the range of the measurements technique to single-digit ppd levels. Characterization of the hygrometer in the integration mode in parallel with an atmospheric pressure ionization mass spectrometer (APIMS) indicated that its accuracy is ±4 ppb, its detection limit is 1 ppd at a 95% confidence limit, and its response time to a 5-ppb moisture intrusion is 50% in 15 minutes.

AB - Study data show that an advanced coulometric hygrometer design incorporating a counterflow sensor cell, a moisture addition feature, and signal integration extends the range of the measurements technique to single-digit ppd levels. Characterization of the hygrometer in the integration mode in parallel with an atmospheric pressure ionization mass spectrometer (APIMS) indicated that its accuracy is ±4 ppb, its detection limit is 1 ppd at a 95% confidence limit, and its response time to a 5-ppb moisture intrusion is 50% in 15 minutes.

UR - http://www.scopus.com/inward/record.url?scp=0029291250&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0029291250&partnerID=8YFLogxK

M3 - Article

AN - SCOPUS:0029291250

VL - 13

SP - 43

EP - 49

JO - MICRO

JF - MICRO

SN - 1081-0595

IS - 4