Evaluation of peripheral nerve stimulators and relationship to possible errors in assessing neuromuscular blockade

K. C. Mylrea, Stuart R Hameroff, J. M. Calkins, C. D. Blitt, L. L. Humphrey

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Voltage and current output characteristics were measured on six commercially available peripheral nerve stimulator devices. The results are evaluated as possible sources of variability in peripheral nerve stimulator function and neuromuscular blockade assessment. The authors found significant differences in output voltage waveform and in maximum current into a 470 ohm load (21.4 to 128 mA.). Output current decreased from 25 to 88% in the different devices, with a load impedance increase from 470 to 10,000 ohms. Due to the variability in peripheral nerve stimulation units and the decrease in current output at higher load impedance, less than supramaximal stimulation is possible with erroneous interpretation of neuromuscular blockade.

Original languageEnglish (US)
Pages (from-to)464-474
Number of pages11
JournalAnesthesiology
Volume60
Issue number5
StatePublished - 1984

Fingerprint

Neuromuscular Blockade
Peripheral Nerves
Electric Impedance
Equipment and Supplies

ASJC Scopus subject areas

  • Anesthesiology and Pain Medicine

Cite this

Evaluation of peripheral nerve stimulators and relationship to possible errors in assessing neuromuscular blockade. / Mylrea, K. C.; Hameroff, Stuart R; Calkins, J. M.; Blitt, C. D.; Humphrey, L. L.

In: Anesthesiology, Vol. 60, No. 5, 1984, p. 464-474.

Research output: Contribution to journalArticle

Mylrea, K. C. ; Hameroff, Stuart R ; Calkins, J. M. ; Blitt, C. D. ; Humphrey, L. L. / Evaluation of peripheral nerve stimulators and relationship to possible errors in assessing neuromuscular blockade. In: Anesthesiology. 1984 ; Vol. 60, No. 5. pp. 464-474.
@article{a9728aa6af974611876b81eb959be919,
title = "Evaluation of peripheral nerve stimulators and relationship to possible errors in assessing neuromuscular blockade",
abstract = "Voltage and current output characteristics were measured on six commercially available peripheral nerve stimulator devices. The results are evaluated as possible sources of variability in peripheral nerve stimulator function and neuromuscular blockade assessment. The authors found significant differences in output voltage waveform and in maximum current into a 470 ohm load (21.4 to 128 mA.). Output current decreased from 25 to 88{\%} in the different devices, with a load impedance increase from 470 to 10,000 ohms. Due to the variability in peripheral nerve stimulation units and the decrease in current output at higher load impedance, less than supramaximal stimulation is possible with erroneous interpretation of neuromuscular blockade.",
author = "Mylrea, {K. C.} and Hameroff, {Stuart R} and Calkins, {J. M.} and Blitt, {C. D.} and Humphrey, {L. L.}",
year = "1984",
language = "English (US)",
volume = "60",
pages = "464--474",
journal = "Anesthesiology",
issn = "0003-3022",
publisher = "Lippincott Williams and Wilkins",
number = "5",

}

TY - JOUR

T1 - Evaluation of peripheral nerve stimulators and relationship to possible errors in assessing neuromuscular blockade

AU - Mylrea, K. C.

AU - Hameroff, Stuart R

AU - Calkins, J. M.

AU - Blitt, C. D.

AU - Humphrey, L. L.

PY - 1984

Y1 - 1984

N2 - Voltage and current output characteristics were measured on six commercially available peripheral nerve stimulator devices. The results are evaluated as possible sources of variability in peripheral nerve stimulator function and neuromuscular blockade assessment. The authors found significant differences in output voltage waveform and in maximum current into a 470 ohm load (21.4 to 128 mA.). Output current decreased from 25 to 88% in the different devices, with a load impedance increase from 470 to 10,000 ohms. Due to the variability in peripheral nerve stimulation units and the decrease in current output at higher load impedance, less than supramaximal stimulation is possible with erroneous interpretation of neuromuscular blockade.

AB - Voltage and current output characteristics were measured on six commercially available peripheral nerve stimulator devices. The results are evaluated as possible sources of variability in peripheral nerve stimulator function and neuromuscular blockade assessment. The authors found significant differences in output voltage waveform and in maximum current into a 470 ohm load (21.4 to 128 mA.). Output current decreased from 25 to 88% in the different devices, with a load impedance increase from 470 to 10,000 ohms. Due to the variability in peripheral nerve stimulation units and the decrease in current output at higher load impedance, less than supramaximal stimulation is possible with erroneous interpretation of neuromuscular blockade.

UR - http://www.scopus.com/inward/record.url?scp=0021319858&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0021319858&partnerID=8YFLogxK

M3 - Article

VL - 60

SP - 464

EP - 474

JO - Anesthesiology

JF - Anesthesiology

SN - 0003-3022

IS - 5

ER -