Interferometric local measurements of high-order aberrations in progressive addition lenses

Sara Chamadoira, Jose M Sasian, Eva Acosta

Research output: Contribution to journalArticle

Abstract

Purpose. The main goal of this work is to show the principles, design, and performance of a modified point diffraction interferometer (PDI) for evaluating the local higher-order aberrations of progressive addition lenses (PALs). Because aberrations have different impacts in visual perception, we propose a device to provide a customized solution with the best permissible amounts and combinations of aberrations for improving a subject's vision. Methods. A PDI has been adapted to measure local high-order aberrations in PALs. Results. High-order aberrations in three different PALs within four relevant circular regions of interest (ROIs) with a radius ranging from 0.4 to 2.4 mm were measured with an accuracy of /10 within the ROI using the tailored PDI. The interferometer also allows for easily choosing the position and number of the ROIs. Conclusions. The interferometric device is compact, robust, and accurate. The operational principle is very simple, and it provides the local high-order aberrations directly without adding additional parts to the interferometer. As expected, the amount of high-order aberrations depends on the chosen ROI of the PAL; the corridor is the more critical region.We found, in accordance with the literature, that, in the corridor, second- and third-order aberrations are dominant and spherical aberration is negligible.

Original languageEnglish (US)
Pages (from-to)1047-1055
Number of pages9
JournalOptometry and Vision Science
Volume92
Issue number11
StatePublished - Oct 27 2015

Fingerprint

Lenses
Equipment and Supplies
Visual Perception

Keywords

  • Local high-order aberration
  • Point diffraction interferometer
  • Progressive addition lenses

ASJC Scopus subject areas

  • Ophthalmology
  • Optometry

Cite this

Interferometric local measurements of high-order aberrations in progressive addition lenses. / Chamadoira, Sara; Sasian, Jose M; Acosta, Eva.

In: Optometry and Vision Science, Vol. 92, No. 11, 27.10.2015, p. 1047-1055.

Research output: Contribution to journalArticle

@article{d7e7250163cc41f6971c2f358819a3e4,
title = "Interferometric local measurements of high-order aberrations in progressive addition lenses",
abstract = "Purpose. The main goal of this work is to show the principles, design, and performance of a modified point diffraction interferometer (PDI) for evaluating the local higher-order aberrations of progressive addition lenses (PALs). Because aberrations have different impacts in visual perception, we propose a device to provide a customized solution with the best permissible amounts and combinations of aberrations for improving a subject's vision. Methods. A PDI has been adapted to measure local high-order aberrations in PALs. Results. High-order aberrations in three different PALs within four relevant circular regions of interest (ROIs) with a radius ranging from 0.4 to 2.4 mm were measured with an accuracy of /10 within the ROI using the tailored PDI. The interferometer also allows for easily choosing the position and number of the ROIs. Conclusions. The interferometric device is compact, robust, and accurate. The operational principle is very simple, and it provides the local high-order aberrations directly without adding additional parts to the interferometer. As expected, the amount of high-order aberrations depends on the chosen ROI of the PAL; the corridor is the more critical region.We found, in accordance with the literature, that, in the corridor, second- and third-order aberrations are dominant and spherical aberration is negligible.",
keywords = "Local high-order aberration, Point diffraction interferometer, Progressive addition lenses",
author = "Sara Chamadoira and Sasian, {Jose M} and Eva Acosta",
year = "2015",
month = "10",
day = "27",
language = "English (US)",
volume = "92",
pages = "1047--1055",
journal = "Optometry and Vision Science",
issn = "1040-5488",
publisher = "Lippincott Williams and Wilkins",
number = "11",

}

TY - JOUR

T1 - Interferometric local measurements of high-order aberrations in progressive addition lenses

AU - Chamadoira, Sara

AU - Sasian, Jose M

AU - Acosta, Eva

PY - 2015/10/27

Y1 - 2015/10/27

N2 - Purpose. The main goal of this work is to show the principles, design, and performance of a modified point diffraction interferometer (PDI) for evaluating the local higher-order aberrations of progressive addition lenses (PALs). Because aberrations have different impacts in visual perception, we propose a device to provide a customized solution with the best permissible amounts and combinations of aberrations for improving a subject's vision. Methods. A PDI has been adapted to measure local high-order aberrations in PALs. Results. High-order aberrations in three different PALs within four relevant circular regions of interest (ROIs) with a radius ranging from 0.4 to 2.4 mm were measured with an accuracy of /10 within the ROI using the tailored PDI. The interferometer also allows for easily choosing the position and number of the ROIs. Conclusions. The interferometric device is compact, robust, and accurate. The operational principle is very simple, and it provides the local high-order aberrations directly without adding additional parts to the interferometer. As expected, the amount of high-order aberrations depends on the chosen ROI of the PAL; the corridor is the more critical region.We found, in accordance with the literature, that, in the corridor, second- and third-order aberrations are dominant and spherical aberration is negligible.

AB - Purpose. The main goal of this work is to show the principles, design, and performance of a modified point diffraction interferometer (PDI) for evaluating the local higher-order aberrations of progressive addition lenses (PALs). Because aberrations have different impacts in visual perception, we propose a device to provide a customized solution with the best permissible amounts and combinations of aberrations for improving a subject's vision. Methods. A PDI has been adapted to measure local high-order aberrations in PALs. Results. High-order aberrations in three different PALs within four relevant circular regions of interest (ROIs) with a radius ranging from 0.4 to 2.4 mm were measured with an accuracy of /10 within the ROI using the tailored PDI. The interferometer also allows for easily choosing the position and number of the ROIs. Conclusions. The interferometric device is compact, robust, and accurate. The operational principle is very simple, and it provides the local high-order aberrations directly without adding additional parts to the interferometer. As expected, the amount of high-order aberrations depends on the chosen ROI of the PAL; the corridor is the more critical region.We found, in accordance with the literature, that, in the corridor, second- and third-order aberrations are dominant and spherical aberration is negligible.

KW - Local high-order aberration

KW - Point diffraction interferometer

KW - Progressive addition lenses

UR - http://www.scopus.com/inward/record.url?scp=84946063105&partnerID=8YFLogxK