Invited paper: Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search

Elizabeth A Krupinski, Hans Roehrig, Jiahua Fan

Research output: Chapter in Book/Report/Conference proceedingConference contribution

1 Citation (Scopus)

Abstract

8 vs 11-bit medical displays were evaluated. Six radiologists viewed 100 chest images. There was no significant difference in ROC Az as a function of bit-depth. Average Az with 8-bits was 0.8284 and with 11-bits was 0.8253. There was a significant difference in viewing time favoring the 11-bit displays.

Original languageEnglish (US)
Title of host publicationDigest of Technical Papers - SID International Symposium
Pages965-966
Number of pages2
Volume38
Edition2
StatePublished - 2007
Event2007 SID International Symposium - Long Beach, CA, United States
Duration: May 23 2007May 25 2007

Other

Other2007 SID International Symposium
CountryUnited States
CityLong Beach, CA
Period5/23/075/25/07

Fingerprint

Display devices

ASJC Scopus subject areas

  • Engineering(all)

Cite this

Krupinski, E. A., Roehrig, H., & Fan, J. (2007). Invited paper: Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search. In Digest of Technical Papers - SID International Symposium (2 ed., Vol. 38, pp. 965-966)

Invited paper : Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search. / Krupinski, Elizabeth A; Roehrig, Hans; Fan, Jiahua.

Digest of Technical Papers - SID International Symposium. Vol. 38 2. ed. 2007. p. 965-966.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Krupinski, EA, Roehrig, H & Fan, J 2007, Invited paper: Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search. in Digest of Technical Papers - SID International Symposium. 2 edn, vol. 38, pp. 965-966, 2007 SID International Symposium, Long Beach, CA, United States, 5/23/07.
Krupinski EA, Roehrig H, Fan J. Invited paper: Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search. In Digest of Technical Papers - SID International Symposium. 2 ed. Vol. 38. 2007. p. 965-966
Krupinski, Elizabeth A ; Roehrig, Hans ; Fan, Jiahua. / Invited paper : Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search. Digest of Technical Papers - SID International Symposium. Vol. 38 2. ed. 2007. pp. 965-966
@inproceedings{ec928bf2b0fc459981dbc5777b239db6,
title = "Invited paper: Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search",
abstract = "8 vs 11-bit medical displays were evaluated. Six radiologists viewed 100 chest images. There was no significant difference in ROC Az as a function of bit-depth. Average Az with 8-bits was 0.8284 and with 11-bits was 0.8253. There was a significant difference in viewing time favoring the 11-bit displays.",
author = "Krupinski, {Elizabeth A} and Hans Roehrig and Jiahua Fan",
year = "2007",
language = "English (US)",
volume = "38",
pages = "965--966",
booktitle = "Digest of Technical Papers - SID International Symposium",
edition = "2",

}

TY - GEN

T1 - Invited paper

T2 - Influence of 8-bit vs 11-bit digital medical displays on observer performance and visual search

AU - Krupinski, Elizabeth A

AU - Roehrig, Hans

AU - Fan, Jiahua

PY - 2007

Y1 - 2007

N2 - 8 vs 11-bit medical displays were evaluated. Six radiologists viewed 100 chest images. There was no significant difference in ROC Az as a function of bit-depth. Average Az with 8-bits was 0.8284 and with 11-bits was 0.8253. There was a significant difference in viewing time favoring the 11-bit displays.

AB - 8 vs 11-bit medical displays were evaluated. Six radiologists viewed 100 chest images. There was no significant difference in ROC Az as a function of bit-depth. Average Az with 8-bits was 0.8284 and with 11-bits was 0.8253. There was a significant difference in viewing time favoring the 11-bit displays.

UR - http://www.scopus.com/inward/record.url?scp=34547932238&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34547932238&partnerID=8YFLogxK

M3 - Conference contribution

AN - SCOPUS:34547932238

VL - 38

SP - 965

EP - 966

BT - Digest of Technical Papers - SID International Symposium

ER -