Laboratory calibration of field reflectance panels

S. F. Biggar, J. Labed, R. P. Santer, P. N. Slater, R. D. Jackson, M. S. Moran

Research output: Contribution to journalArticlepeer-review

55 Scopus citations

Abstract

A method used for calibrating field reflectance panels in the visible and shortwave infrared wavelength range is described. The directional reflectance factor of painted barium sulfate (BaSO4) panels is determined. The reference for this method is the hemispherical reflectance of pressed polytetrafluoroethylene (halon) powder prepared according to National Bureau of Standards (NBS) directions. The panels and a radiometer are mounted on rotation stages to measure the reflectance factor at different incidence and view angles. The sensor can be any laboratory or field filter radiometer small enough to mount on the apparatus. The method is used to measure the reflectance factors of halon and BaSO4 panels between 0.45 and 0.85 micrometers. These reflectance factors are compared to those measured by a field apparatus. The results agree to within 0.013 in reflectance at incidence angles between 15 and 75 degrees.

Original languageEnglish (US)
Pages (from-to)232-240
Number of pages9
JournalProceedings of SPIE - The International Society for Optical Engineering
Volume924
DOIs
StatePublished - Oct 12 1988

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Condensed Matter Physics
  • Computer Science Applications
  • Applied Mathematics
  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Laboratory calibration of field reflectance panels'. Together they form a unique fingerprint.

Cite this