Large-vessel distensibility measurement with electrocardiographically gated multidetector CT: Phantom study and initial experience

Jie Zhang, Joel G. Fletcher, Terri J. Vrtiska, Armando Manduca, Jess L Thompson, Madhavan L. Raghavan, Robert J. Wentz, Cynthia H. McCollough

Research output: Contribution to journalArticle

22 Citations (Scopus)

Abstract

The purpose of this study was to prospectively examine vessel distensibility measurements by using electrocardiographically gated multidetector computed tomography (CT) in a phantom compared with measurements by using a digital camera and to examine feasibility in humans. Large-vessel phantoms were constructed, using a pulsatile flow pump, and absolute diameter and percentage diameter changes during pulsation were measured. After institutional review board approval and patient consent were obtained, the abdominal aorta of four patients was scanned with an electrocardiographically gated CT protocol. The mean difference in percentage diameter change between CT and optical measurements for the phantom ranged from -0.47% to 0.14%. The range of area changes in five locations along the abdominal aorta in four patients was 2.97%-37.16%. Findings of this study indicate that electrocardiographically gated CT angiography data reconstructed across cardiac phases permit measurements of large-vessel distensibility in a phantom model and that vessel distensibility measurement in humans may be possible.

Original languageEnglish (US)
Pages (from-to)258-266
Number of pages9
JournalRadiology
Volume245
Issue number1
DOIs
StatePublished - Oct 2007
Externally publishedYes

Fingerprint

Multidetector Computed Tomography
Abdominal Aorta
Optical Tomography
Pulsatile Flow
Research Ethics Committees
Tomography

ASJC Scopus subject areas

  • Radiological and Ultrasound Technology

Cite this

Large-vessel distensibility measurement with electrocardiographically gated multidetector CT : Phantom study and initial experience. / Zhang, Jie; Fletcher, Joel G.; Vrtiska, Terri J.; Manduca, Armando; Thompson, Jess L; Raghavan, Madhavan L.; Wentz, Robert J.; McCollough, Cynthia H.

In: Radiology, Vol. 245, No. 1, 10.2007, p. 258-266.

Research output: Contribution to journalArticle

Zhang, J, Fletcher, JG, Vrtiska, TJ, Manduca, A, Thompson, JL, Raghavan, ML, Wentz, RJ & McCollough, CH 2007, 'Large-vessel distensibility measurement with electrocardiographically gated multidetector CT: Phantom study and initial experience', Radiology, vol. 245, no. 1, pp. 258-266. https://doi.org/10.1148/radiol.2451060530
Zhang, Jie ; Fletcher, Joel G. ; Vrtiska, Terri J. ; Manduca, Armando ; Thompson, Jess L ; Raghavan, Madhavan L. ; Wentz, Robert J. ; McCollough, Cynthia H. / Large-vessel distensibility measurement with electrocardiographically gated multidetector CT : Phantom study and initial experience. In: Radiology. 2007 ; Vol. 245, No. 1. pp. 258-266.
@article{260a912e62ba4f40aec2443616d65fc9,
title = "Large-vessel distensibility measurement with electrocardiographically gated multidetector CT: Phantom study and initial experience",
abstract = "The purpose of this study was to prospectively examine vessel distensibility measurements by using electrocardiographically gated multidetector computed tomography (CT) in a phantom compared with measurements by using a digital camera and to examine feasibility in humans. Large-vessel phantoms were constructed, using a pulsatile flow pump, and absolute diameter and percentage diameter changes during pulsation were measured. After institutional review board approval and patient consent were obtained, the abdominal aorta of four patients was scanned with an electrocardiographically gated CT protocol. The mean difference in percentage diameter change between CT and optical measurements for the phantom ranged from -0.47{\%} to 0.14{\%}. The range of area changes in five locations along the abdominal aorta in four patients was 2.97{\%}-37.16{\%}. Findings of this study indicate that electrocardiographically gated CT angiography data reconstructed across cardiac phases permit measurements of large-vessel distensibility in a phantom model and that vessel distensibility measurement in humans may be possible.",
author = "Jie Zhang and Fletcher, {Joel G.} and Vrtiska, {Terri J.} and Armando Manduca and Thompson, {Jess L} and Raghavan, {Madhavan L.} and Wentz, {Robert J.} and McCollough, {Cynthia H.}",
year = "2007",
month = "10",
doi = "10.1148/radiol.2451060530",
language = "English (US)",
volume = "245",
pages = "258--266",
journal = "Radiology",
issn = "0033-8419",
publisher = "Radiological Society of North America Inc.",
number = "1",

}

TY - JOUR

T1 - Large-vessel distensibility measurement with electrocardiographically gated multidetector CT

T2 - Phantom study and initial experience

AU - Zhang, Jie

AU - Fletcher, Joel G.

AU - Vrtiska, Terri J.

AU - Manduca, Armando

AU - Thompson, Jess L

AU - Raghavan, Madhavan L.

AU - Wentz, Robert J.

AU - McCollough, Cynthia H.

PY - 2007/10

Y1 - 2007/10

N2 - The purpose of this study was to prospectively examine vessel distensibility measurements by using electrocardiographically gated multidetector computed tomography (CT) in a phantom compared with measurements by using a digital camera and to examine feasibility in humans. Large-vessel phantoms were constructed, using a pulsatile flow pump, and absolute diameter and percentage diameter changes during pulsation were measured. After institutional review board approval and patient consent were obtained, the abdominal aorta of four patients was scanned with an electrocardiographically gated CT protocol. The mean difference in percentage diameter change between CT and optical measurements for the phantom ranged from -0.47% to 0.14%. The range of area changes in five locations along the abdominal aorta in four patients was 2.97%-37.16%. Findings of this study indicate that electrocardiographically gated CT angiography data reconstructed across cardiac phases permit measurements of large-vessel distensibility in a phantom model and that vessel distensibility measurement in humans may be possible.

AB - The purpose of this study was to prospectively examine vessel distensibility measurements by using electrocardiographically gated multidetector computed tomography (CT) in a phantom compared with measurements by using a digital camera and to examine feasibility in humans. Large-vessel phantoms were constructed, using a pulsatile flow pump, and absolute diameter and percentage diameter changes during pulsation were measured. After institutional review board approval and patient consent were obtained, the abdominal aorta of four patients was scanned with an electrocardiographically gated CT protocol. The mean difference in percentage diameter change between CT and optical measurements for the phantom ranged from -0.47% to 0.14%. The range of area changes in five locations along the abdominal aorta in four patients was 2.97%-37.16%. Findings of this study indicate that electrocardiographically gated CT angiography data reconstructed across cardiac phases permit measurements of large-vessel distensibility in a phantom model and that vessel distensibility measurement in humans may be possible.

UR - http://www.scopus.com/inward/record.url?scp=34548776354&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=34548776354&partnerID=8YFLogxK

U2 - 10.1148/radiol.2451060530

DO - 10.1148/radiol.2451060530

M3 - Article

C2 - 17885194

AN - SCOPUS:34548776354

VL - 245

SP - 258

EP - 266

JO - Radiology

JF - Radiology

SN - 0033-8419

IS - 1