Maximum-likelihood calibration of an x-ray computed tomography system

Jared W. Moore, Roel Van Holen, Harrison H Barrett, Lars R Furenlid

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

We present a maximum-likelihood (ML) method for calibrating the geometrical parameters of an x-ray computed tomography (CT) system. This method makes use of the full image data and not a reduced set of data. This algorithm is particularly useful for CT systems that change their geometry during the CT acquisition, such as an adaptive CT scan. Our ML search method uses a contracting-grid algorithm that does not require initial starting values to perform its estimate, thus avoiding problems associated with choosing initialization values.

Original languageEnglish (US)
Title of host publicationIEEE Nuclear Science Symposium Conference Record
Pages2614-2616
Number of pages3
DOIs
StatePublished - 2010
Event2010 IEEE Nuclear Science Symposium, Medical Imaging Conference, NSS/MIC 2010 and 17th International Workshop on Room-Temperature Semiconductor X-ray and Gamma-ray Detectors, RTSD 2010 - Knoxville, TN, United States
Duration: Oct 30 2010Nov 6 2010

Other

Other2010 IEEE Nuclear Science Symposium, Medical Imaging Conference, NSS/MIC 2010 and 17th International Workshop on Room-Temperature Semiconductor X-ray and Gamma-ray Detectors, RTSD 2010
CountryUnited States
CityKnoxville, TN
Period10/30/1011/6/10

Fingerprint

Calibration
tomography
Tomography
X-Rays
x rays
calibrating
acquisition
grids
estimates
geometry

ASJC Scopus subject areas

  • Radiation
  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

Cite this

Moore, J. W., Van Holen, R., Barrett, H. H., & Furenlid, L. R. (2010). Maximum-likelihood calibration of an x-ray computed tomography system. In IEEE Nuclear Science Symposium Conference Record (pp. 2614-2616). [5874262] https://doi.org/10.1109/NSSMIC.2010.5874262

Maximum-likelihood calibration of an x-ray computed tomography system. / Moore, Jared W.; Van Holen, Roel; Barrett, Harrison H; Furenlid, Lars R.

IEEE Nuclear Science Symposium Conference Record. 2010. p. 2614-2616 5874262.

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Moore, JW, Van Holen, R, Barrett, HH & Furenlid, LR 2010, Maximum-likelihood calibration of an x-ray computed tomography system. in IEEE Nuclear Science Symposium Conference Record., 5874262, pp. 2614-2616, 2010 IEEE Nuclear Science Symposium, Medical Imaging Conference, NSS/MIC 2010 and 17th International Workshop on Room-Temperature Semiconductor X-ray and Gamma-ray Detectors, RTSD 2010, Knoxville, TN, United States, 10/30/10. https://doi.org/10.1109/NSSMIC.2010.5874262
Moore JW, Van Holen R, Barrett HH, Furenlid LR. Maximum-likelihood calibration of an x-ray computed tomography system. In IEEE Nuclear Science Symposium Conference Record. 2010. p. 2614-2616. 5874262 https://doi.org/10.1109/NSSMIC.2010.5874262
Moore, Jared W. ; Van Holen, Roel ; Barrett, Harrison H ; Furenlid, Lars R. / Maximum-likelihood calibration of an x-ray computed tomography system. IEEE Nuclear Science Symposium Conference Record. 2010. pp. 2614-2616
@inproceedings{ba91085391d746e2a58a16378eb6ae17,
title = "Maximum-likelihood calibration of an x-ray computed tomography system",
abstract = "We present a maximum-likelihood (ML) method for calibrating the geometrical parameters of an x-ray computed tomography (CT) system. This method makes use of the full image data and not a reduced set of data. This algorithm is particularly useful for CT systems that change their geometry during the CT acquisition, such as an adaptive CT scan. Our ML search method uses a contracting-grid algorithm that does not require initial starting values to perform its estimate, thus avoiding problems associated with choosing initialization values.",
author = "Moore, {Jared W.} and {Van Holen}, Roel and Barrett, {Harrison H} and Furenlid, {Lars R}",
year = "2010",
doi = "10.1109/NSSMIC.2010.5874262",
language = "English (US)",
isbn = "9781424491063",
pages = "2614--2616",
booktitle = "IEEE Nuclear Science Symposium Conference Record",

}

TY - GEN

T1 - Maximum-likelihood calibration of an x-ray computed tomography system

AU - Moore, Jared W.

AU - Van Holen, Roel

AU - Barrett, Harrison H

AU - Furenlid, Lars R

PY - 2010

Y1 - 2010

N2 - We present a maximum-likelihood (ML) method for calibrating the geometrical parameters of an x-ray computed tomography (CT) system. This method makes use of the full image data and not a reduced set of data. This algorithm is particularly useful for CT systems that change their geometry during the CT acquisition, such as an adaptive CT scan. Our ML search method uses a contracting-grid algorithm that does not require initial starting values to perform its estimate, thus avoiding problems associated with choosing initialization values.

AB - We present a maximum-likelihood (ML) method for calibrating the geometrical parameters of an x-ray computed tomography (CT) system. This method makes use of the full image data and not a reduced set of data. This algorithm is particularly useful for CT systems that change their geometry during the CT acquisition, such as an adaptive CT scan. Our ML search method uses a contracting-grid algorithm that does not require initial starting values to perform its estimate, thus avoiding problems associated with choosing initialization values.

UR - http://www.scopus.com/inward/record.url?scp=79960297251&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=79960297251&partnerID=8YFLogxK

U2 - 10.1109/NSSMIC.2010.5874262

DO - 10.1109/NSSMIC.2010.5874262

M3 - Conference contribution

AN - SCOPUS:79960297251

SN - 9781424491063

SP - 2614

EP - 2616

BT - IEEE Nuclear Science Symposium Conference Record