Novel 3D computer-automated threshold Amsler grid visual field testing of scotomas in patients with glaucoma

Dieuthu T. Nguyen, Ali Fahimi, Wolfgang Fink, Paul P. Nazemi, Janet K. Kim, Alfredo A. Sadun

Research output: Contribution to journalArticle

8 Citations (Scopus)

Abstract

Purpose. Three-dimensional (3D) computer-automated threshold Amsler grid testing was used to identify and characterize scotomas typical of glaucoma. Methods. The 3D test exhibits a grid on a computer screen at a preselected grayscale and angular resolution, and allows patients to trace those areas on the grid that are missing in their visual field using a touch screen. Eleven eyes in patients with an established diagnosis of glaucoma were examined according to the above protocol. A total of 23 eyes from normal subjects were used as controls. The 5-minute test required that patients repeatedly outline scotomas on a touch screen with varied displays of contrast while maintaining their gaze on a central fixation marker. A 3D depiction of the visual field defects was then obtained that was further characterized by the location, shape, extent, depth, and slope of the scotomas. Results. In this pilot study, the 3D depiction of visual field loss demonstrated paracentral, superior and inferior altitudinal, and nasal step defects consistent with glaucomatous damage. The 3D depiction showed a shape, extent, depth, and slope that are consistent with the severity of damage. Conclusions. The 3D test identified and characterized scotomas typical of glaucoma. The test provides several advantages over conventional perimetry including additional information through 3D depiction of scotomas with the addition of contrast sensitivity and a higher angular/ spatial resolution. Improved patient compliance and reliability through shorter testing time and potential interactive accessibility and distribution over the Internet further characterize the test.

Original languageEnglish (US)
Pages (from-to)776-782
Number of pages7
JournalEuropean Journal of Ophthalmology
Volume19
Issue number5
StatePublished - 2009
Externally publishedYes

Fingerprint

Scotoma
Visual Fields
Glaucoma
Visual Field Tests
Contrast Sensitivity
Patient Compliance
Nose
Internet

Keywords

  • Glaucoma
  • Glaucomatous visual field defects
  • Perimetry
  • Screening
  • Visual field

ASJC Scopus subject areas

  • Ophthalmology

Cite this

Novel 3D computer-automated threshold Amsler grid visual field testing of scotomas in patients with glaucoma. / Nguyen, Dieuthu T.; Fahimi, Ali; Fink, Wolfgang; Nazemi, Paul P.; Kim, Janet K.; Sadun, Alfredo A.

In: European Journal of Ophthalmology, Vol. 19, No. 5, 2009, p. 776-782.

Research output: Contribution to journalArticle

Nguyen, Dieuthu T. ; Fahimi, Ali ; Fink, Wolfgang ; Nazemi, Paul P. ; Kim, Janet K. ; Sadun, Alfredo A. / Novel 3D computer-automated threshold Amsler grid visual field testing of scotomas in patients with glaucoma. In: European Journal of Ophthalmology. 2009 ; Vol. 19, No. 5. pp. 776-782.
@article{fc62a631c20a405fac7dfa1a37809e76,
title = "Novel 3D computer-automated threshold Amsler grid visual field testing of scotomas in patients with glaucoma",
abstract = "Purpose. Three-dimensional (3D) computer-automated threshold Amsler grid testing was used to identify and characterize scotomas typical of glaucoma. Methods. The 3D test exhibits a grid on a computer screen at a preselected grayscale and angular resolution, and allows patients to trace those areas on the grid that are missing in their visual field using a touch screen. Eleven eyes in patients with an established diagnosis of glaucoma were examined according to the above protocol. A total of 23 eyes from normal subjects were used as controls. The 5-minute test required that patients repeatedly outline scotomas on a touch screen with varied displays of contrast while maintaining their gaze on a central fixation marker. A 3D depiction of the visual field defects was then obtained that was further characterized by the location, shape, extent, depth, and slope of the scotomas. Results. In this pilot study, the 3D depiction of visual field loss demonstrated paracentral, superior and inferior altitudinal, and nasal step defects consistent with glaucomatous damage. The 3D depiction showed a shape, extent, depth, and slope that are consistent with the severity of damage. Conclusions. The 3D test identified and characterized scotomas typical of glaucoma. The test provides several advantages over conventional perimetry including additional information through 3D depiction of scotomas with the addition of contrast sensitivity and a higher angular/ spatial resolution. Improved patient compliance and reliability through shorter testing time and potential interactive accessibility and distribution over the Internet further characterize the test.",
keywords = "Glaucoma, Glaucomatous visual field defects, Perimetry, Screening, Visual field",
author = "Nguyen, {Dieuthu T.} and Ali Fahimi and Wolfgang Fink and Nazemi, {Paul P.} and Kim, {Janet K.} and Sadun, {Alfredo A.}",
year = "2009",
language = "English (US)",
volume = "19",
pages = "776--782",
journal = "European Journal of Ophthalmology",
issn = "1120-6721",
publisher = "Wichtig Publishing",
number = "5",

}

TY - JOUR

T1 - Novel 3D computer-automated threshold Amsler grid visual field testing of scotomas in patients with glaucoma

AU - Nguyen, Dieuthu T.

AU - Fahimi, Ali

AU - Fink, Wolfgang

AU - Nazemi, Paul P.

AU - Kim, Janet K.

AU - Sadun, Alfredo A.

PY - 2009

Y1 - 2009

N2 - Purpose. Three-dimensional (3D) computer-automated threshold Amsler grid testing was used to identify and characterize scotomas typical of glaucoma. Methods. The 3D test exhibits a grid on a computer screen at a preselected grayscale and angular resolution, and allows patients to trace those areas on the grid that are missing in their visual field using a touch screen. Eleven eyes in patients with an established diagnosis of glaucoma were examined according to the above protocol. A total of 23 eyes from normal subjects were used as controls. The 5-minute test required that patients repeatedly outline scotomas on a touch screen with varied displays of contrast while maintaining their gaze on a central fixation marker. A 3D depiction of the visual field defects was then obtained that was further characterized by the location, shape, extent, depth, and slope of the scotomas. Results. In this pilot study, the 3D depiction of visual field loss demonstrated paracentral, superior and inferior altitudinal, and nasal step defects consistent with glaucomatous damage. The 3D depiction showed a shape, extent, depth, and slope that are consistent with the severity of damage. Conclusions. The 3D test identified and characterized scotomas typical of glaucoma. The test provides several advantages over conventional perimetry including additional information through 3D depiction of scotomas with the addition of contrast sensitivity and a higher angular/ spatial resolution. Improved patient compliance and reliability through shorter testing time and potential interactive accessibility and distribution over the Internet further characterize the test.

AB - Purpose. Three-dimensional (3D) computer-automated threshold Amsler grid testing was used to identify and characterize scotomas typical of glaucoma. Methods. The 3D test exhibits a grid on a computer screen at a preselected grayscale and angular resolution, and allows patients to trace those areas on the grid that are missing in their visual field using a touch screen. Eleven eyes in patients with an established diagnosis of glaucoma were examined according to the above protocol. A total of 23 eyes from normal subjects were used as controls. The 5-minute test required that patients repeatedly outline scotomas on a touch screen with varied displays of contrast while maintaining their gaze on a central fixation marker. A 3D depiction of the visual field defects was then obtained that was further characterized by the location, shape, extent, depth, and slope of the scotomas. Results. In this pilot study, the 3D depiction of visual field loss demonstrated paracentral, superior and inferior altitudinal, and nasal step defects consistent with glaucomatous damage. The 3D depiction showed a shape, extent, depth, and slope that are consistent with the severity of damage. Conclusions. The 3D test identified and characterized scotomas typical of glaucoma. The test provides several advantages over conventional perimetry including additional information through 3D depiction of scotomas with the addition of contrast sensitivity and a higher angular/ spatial resolution. Improved patient compliance and reliability through shorter testing time and potential interactive accessibility and distribution over the Internet further characterize the test.

KW - Glaucoma

KW - Glaucomatous visual field defects

KW - Perimetry

KW - Screening

KW - Visual field

UR - http://www.scopus.com/inward/record.url?scp=70549089036&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=70549089036&partnerID=8YFLogxK

M3 - Article

VL - 19

SP - 776

EP - 782

JO - European Journal of Ophthalmology

JF - European Journal of Ophthalmology

SN - 1120-6721

IS - 5

ER -