The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output

D. C. Fisher, D. J. Sahn, Mark J Friedman, Douglas F Larson, L. M. Valdes-Cruz, S. Horowitz, S. J. Goldberg, H. D. Allen

Research output: Contribution to journalArticle

132 Citations (Scopus)

Abstract

We developed and validated a mitral valve orifice method for Doppler cardiac output determination. In 15 open-chest dogs, cardiac output was controlled and measured by a roller pump interposed between the right atrium and pulmonary artery as a right-heart bypass. Left hearts flows were measured in the open-chest dog model by Doppler measurements at the mitral valve orifice and compared not only to volume flow measured by the roller pump, but to electromagnetic flow meters as well. The maximum mitral valve orifice area was measured off short-axis two-dimensional echocardiographic views by planimetry. The maximal orifice was then adjusted for its diastolic variation in size by calculating a ratio of mean-to-maximal mitral valve separation on a derived M-mode echocardiogram. Flow was sampled parallel to mitral valve inflow in a four-chamber plane. The multiplication of mean flow throughout the output determination that could be compared to the roller pump measurement. Fifty-two cardiac output determinations over roller pump values of 1-5 l/min yielded a high correlation between roller pump flows and Doppler (r = 0.97 ± 0.23 l/min). Our study shows that the mitral valve orifice provides an accurate site for Doppler cardiac output measurements.

Original languageEnglish (US)
Pages (from-to)872-877
Number of pages6
JournalCirculation
Volume67
Issue number4
StatePublished - 1983

Fingerprint

Mitral Valve
Cardiac Output
Thorax
Right Heart Bypass
Dogs
Electromagnetic Phenomena
Heart Atria
Pulmonary Artery

ASJC Scopus subject areas

  • Physiology
  • Cardiology and Cardiovascular Medicine

Cite this

Fisher, D. C., Sahn, D. J., Friedman, M. J., Larson, D. F., Valdes-Cruz, L. M., Horowitz, S., ... Allen, H. D. (1983). The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output. Circulation, 67(4), 872-877.

The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output. / Fisher, D. C.; Sahn, D. J.; Friedman, Mark J; Larson, Douglas F; Valdes-Cruz, L. M.; Horowitz, S.; Goldberg, S. J.; Allen, H. D.

In: Circulation, Vol. 67, No. 4, 1983, p. 872-877.

Research output: Contribution to journalArticle

Fisher, DC, Sahn, DJ, Friedman, MJ, Larson, DF, Valdes-Cruz, LM, Horowitz, S, Goldberg, SJ & Allen, HD 1983, 'The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output', Circulation, vol. 67, no. 4, pp. 872-877.
Fisher, D. C. ; Sahn, D. J. ; Friedman, Mark J ; Larson, Douglas F ; Valdes-Cruz, L. M. ; Horowitz, S. ; Goldberg, S. J. ; Allen, H. D. / The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output. In: Circulation. 1983 ; Vol. 67, No. 4. pp. 872-877.
@article{041c3a4ead8e415aa2e37795db91aa61,
title = "The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output",
abstract = "We developed and validated a mitral valve orifice method for Doppler cardiac output determination. In 15 open-chest dogs, cardiac output was controlled and measured by a roller pump interposed between the right atrium and pulmonary artery as a right-heart bypass. Left hearts flows were measured in the open-chest dog model by Doppler measurements at the mitral valve orifice and compared not only to volume flow measured by the roller pump, but to electromagnetic flow meters as well. The maximum mitral valve orifice area was measured off short-axis two-dimensional echocardiographic views by planimetry. The maximal orifice was then adjusted for its diastolic variation in size by calculating a ratio of mean-to-maximal mitral valve separation on a derived M-mode echocardiogram. Flow was sampled parallel to mitral valve inflow in a four-chamber plane. The multiplication of mean flow throughout the output determination that could be compared to the roller pump measurement. Fifty-two cardiac output determinations over roller pump values of 1-5 l/min yielded a high correlation between roller pump flows and Doppler (r = 0.97 ± 0.23 l/min). Our study shows that the mitral valve orifice provides an accurate site for Doppler cardiac output measurements.",
author = "Fisher, {D. C.} and Sahn, {D. J.} and Friedman, {Mark J} and Larson, {Douglas F} and Valdes-Cruz, {L. M.} and S. Horowitz and Goldberg, {S. J.} and Allen, {H. D.}",
year = "1983",
language = "English (US)",
volume = "67",
pages = "872--877",
journal = "Circulation",
issn = "0009-7322",
publisher = "Lippincott Williams and Wilkins",
number = "4",

}

TY - JOUR

T1 - The mitral valve orifice method for noninvasive two-dimensional echo Doppler determinations of cardiac output

AU - Fisher, D. C.

AU - Sahn, D. J.

AU - Friedman, Mark J

AU - Larson, Douglas F

AU - Valdes-Cruz, L. M.

AU - Horowitz, S.

AU - Goldberg, S. J.

AU - Allen, H. D.

PY - 1983

Y1 - 1983

N2 - We developed and validated a mitral valve orifice method for Doppler cardiac output determination. In 15 open-chest dogs, cardiac output was controlled and measured by a roller pump interposed between the right atrium and pulmonary artery as a right-heart bypass. Left hearts flows were measured in the open-chest dog model by Doppler measurements at the mitral valve orifice and compared not only to volume flow measured by the roller pump, but to electromagnetic flow meters as well. The maximum mitral valve orifice area was measured off short-axis two-dimensional echocardiographic views by planimetry. The maximal orifice was then adjusted for its diastolic variation in size by calculating a ratio of mean-to-maximal mitral valve separation on a derived M-mode echocardiogram. Flow was sampled parallel to mitral valve inflow in a four-chamber plane. The multiplication of mean flow throughout the output determination that could be compared to the roller pump measurement. Fifty-two cardiac output determinations over roller pump values of 1-5 l/min yielded a high correlation between roller pump flows and Doppler (r = 0.97 ± 0.23 l/min). Our study shows that the mitral valve orifice provides an accurate site for Doppler cardiac output measurements.

AB - We developed and validated a mitral valve orifice method for Doppler cardiac output determination. In 15 open-chest dogs, cardiac output was controlled and measured by a roller pump interposed between the right atrium and pulmonary artery as a right-heart bypass. Left hearts flows were measured in the open-chest dog model by Doppler measurements at the mitral valve orifice and compared not only to volume flow measured by the roller pump, but to electromagnetic flow meters as well. The maximum mitral valve orifice area was measured off short-axis two-dimensional echocardiographic views by planimetry. The maximal orifice was then adjusted for its diastolic variation in size by calculating a ratio of mean-to-maximal mitral valve separation on a derived M-mode echocardiogram. Flow was sampled parallel to mitral valve inflow in a four-chamber plane. The multiplication of mean flow throughout the output determination that could be compared to the roller pump measurement. Fifty-two cardiac output determinations over roller pump values of 1-5 l/min yielded a high correlation between roller pump flows and Doppler (r = 0.97 ± 0.23 l/min). Our study shows that the mitral valve orifice provides an accurate site for Doppler cardiac output measurements.

UR - http://www.scopus.com/inward/record.url?scp=0020665265&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=0020665265&partnerID=8YFLogxK

M3 - Article

VL - 67

SP - 872

EP - 877

JO - Circulation

JF - Circulation

SN - 0009-7322

IS - 4

ER -