The speech perception gap in cochlear implant patients

Zechariah G. Franks, Abraham Jacob

Research output: Contribution to journalArticle

Abstract

Objective: To determine how commonly word recognition scores obtained using insert microphones (PB max) overestimate word recognition scores obtained through appropriately fit hearing aids (A-WRS). Methods: Aided speech recognition tests may not be performed during routine hearing aid fittings; however, they are regularly performed for cochlear implant (CI) candidacy evaluation. Therefore, audiologic data from CI recipients were queried in a retrospective cohort study at a tertiary care center. PB max and A-WRS were obtained. The Speech Perception (SP) gap, defined as PB max minus A-WRS, was calculated for each patient and a high SP gap was defined as ≥20%. Results: Analyzing 78 patients with complete data, 30 patients had PB max ≥20%. Of these, 18 (60%) had a high SP gap. Eighteen of the 78 patients had PB max ≥40%, and of these patients, 15 (83%) had a high SP gap. Discussion/conclusion: A Speech Perception Gap of ≥20% may exist in a sizable percentage of patients with hearing loss. Our pilot study suggests that over 80% of these patients could have Class D hearing (speech recognition <50%) using conventional hearing aids and may be better served using alternate rehabilitation strategies such as middle ear or cochlear implants. Therefore, aided speech testing should be performed as part of a verified hearing aid fit in all patients, especially those with PB Max ranging from 40% to 70%.

Original languageEnglish (US)
JournalCochlear implants international
DOIs
StatePublished - Jan 1 2019
Externally publishedYes

Fingerprint

Speech Perception
Cochlear Implants
Hearing Aids
Ossicular Prosthesis
Hearing Loss
Tertiary Care Centers
Hearing
Cohort Studies
Rehabilitation
Retrospective Studies
Recognition (Psychology)

Keywords

  • Cochlear Implant
  • Hearing Aid
  • Hearing Aid Fit
  • Hearing Aid Fitting
  • Sensorineural Hearing Loss
  • Speech Perception Gap
  • Speech Recognition
  • Word Recognition Score

ASJC Scopus subject areas

  • Otorhinolaryngology
  • Speech and Hearing

Cite this

The speech perception gap in cochlear implant patients. / Franks, Zechariah G.; Jacob, Abraham.

In: Cochlear implants international, 01.01.2019.

Research output: Contribution to journalArticle

@article{bd70e1c9db80477290cdaa882aae866b,
title = "The speech perception gap in cochlear implant patients",
abstract = "Objective: To determine how commonly word recognition scores obtained using insert microphones (PB max) overestimate word recognition scores obtained through appropriately fit hearing aids (A-WRS). Methods: Aided speech recognition tests may not be performed during routine hearing aid fittings; however, they are regularly performed for cochlear implant (CI) candidacy evaluation. Therefore, audiologic data from CI recipients were queried in a retrospective cohort study at a tertiary care center. PB max and A-WRS were obtained. The Speech Perception (SP) gap, defined as PB max minus A-WRS, was calculated for each patient and a high SP gap was defined as ≥20{\%}. Results: Analyzing 78 patients with complete data, 30 patients had PB max ≥20{\%}. Of these, 18 (60{\%}) had a high SP gap. Eighteen of the 78 patients had PB max ≥40{\%}, and of these patients, 15 (83{\%}) had a high SP gap. Discussion/conclusion: A Speech Perception Gap of ≥20{\%} may exist in a sizable percentage of patients with hearing loss. Our pilot study suggests that over 80{\%} of these patients could have Class D hearing (speech recognition <50{\%}) using conventional hearing aids and may be better served using alternate rehabilitation strategies such as middle ear or cochlear implants. Therefore, aided speech testing should be performed as part of a verified hearing aid fit in all patients, especially those with PB Max ranging from 40{\%} to 70{\%}.",
keywords = "Cochlear Implant, Hearing Aid, Hearing Aid Fit, Hearing Aid Fitting, Sensorineural Hearing Loss, Speech Perception Gap, Speech Recognition, Word Recognition Score",
author = "Franks, {Zechariah G.} and Abraham Jacob",
year = "2019",
month = "1",
day = "1",
doi = "10.1080/14670100.2019.1582165",
language = "English (US)",
journal = "Cochlear Implants International",
issn = "1467-0100",
publisher = "Maney Publishing",

}

TY - JOUR

T1 - The speech perception gap in cochlear implant patients

AU - Franks, Zechariah G.

AU - Jacob, Abraham

PY - 2019/1/1

Y1 - 2019/1/1

N2 - Objective: To determine how commonly word recognition scores obtained using insert microphones (PB max) overestimate word recognition scores obtained through appropriately fit hearing aids (A-WRS). Methods: Aided speech recognition tests may not be performed during routine hearing aid fittings; however, they are regularly performed for cochlear implant (CI) candidacy evaluation. Therefore, audiologic data from CI recipients were queried in a retrospective cohort study at a tertiary care center. PB max and A-WRS were obtained. The Speech Perception (SP) gap, defined as PB max minus A-WRS, was calculated for each patient and a high SP gap was defined as ≥20%. Results: Analyzing 78 patients with complete data, 30 patients had PB max ≥20%. Of these, 18 (60%) had a high SP gap. Eighteen of the 78 patients had PB max ≥40%, and of these patients, 15 (83%) had a high SP gap. Discussion/conclusion: A Speech Perception Gap of ≥20% may exist in a sizable percentage of patients with hearing loss. Our pilot study suggests that over 80% of these patients could have Class D hearing (speech recognition <50%) using conventional hearing aids and may be better served using alternate rehabilitation strategies such as middle ear or cochlear implants. Therefore, aided speech testing should be performed as part of a verified hearing aid fit in all patients, especially those with PB Max ranging from 40% to 70%.

AB - Objective: To determine how commonly word recognition scores obtained using insert microphones (PB max) overestimate word recognition scores obtained through appropriately fit hearing aids (A-WRS). Methods: Aided speech recognition tests may not be performed during routine hearing aid fittings; however, they are regularly performed for cochlear implant (CI) candidacy evaluation. Therefore, audiologic data from CI recipients were queried in a retrospective cohort study at a tertiary care center. PB max and A-WRS were obtained. The Speech Perception (SP) gap, defined as PB max minus A-WRS, was calculated for each patient and a high SP gap was defined as ≥20%. Results: Analyzing 78 patients with complete data, 30 patients had PB max ≥20%. Of these, 18 (60%) had a high SP gap. Eighteen of the 78 patients had PB max ≥40%, and of these patients, 15 (83%) had a high SP gap. Discussion/conclusion: A Speech Perception Gap of ≥20% may exist in a sizable percentage of patients with hearing loss. Our pilot study suggests that over 80% of these patients could have Class D hearing (speech recognition <50%) using conventional hearing aids and may be better served using alternate rehabilitation strategies such as middle ear or cochlear implants. Therefore, aided speech testing should be performed as part of a verified hearing aid fit in all patients, especially those with PB Max ranging from 40% to 70%.

KW - Cochlear Implant

KW - Hearing Aid

KW - Hearing Aid Fit

KW - Hearing Aid Fitting

KW - Sensorineural Hearing Loss

KW - Speech Perception Gap

KW - Speech Recognition

KW - Word Recognition Score

UR - http://www.scopus.com/inward/record.url?scp=85062355565&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85062355565&partnerID=8YFLogxK

U2 - 10.1080/14670100.2019.1582165

DO - 10.1080/14670100.2019.1582165

M3 - Article

C2 - 30806176

AN - SCOPUS:85062355565

JO - Cochlear Implants International

JF - Cochlear Implants International

SN - 1467-0100

ER -