Ultrasound Elasticity Imaging for Determining the Mechanical Properties of Human Posterior Tibial Tendon: A Cadaveric Study

Liang Gao, Justin S. Yuan, Gregory J. Heden, John Szivek, Mihra Taljanovic, Daniel L Latt, Russell S Witte

Research output: Contribution to journalArticle

9 Citations (Scopus)

Abstract

Posterior tibial tendon dysfunction (PTTD) is a common degenerative condition leading to a severe impairment of gait. There is currently no effective method to determine whether a patient with advanced PTTD would benefit from several months of bracing and physical therapy or ultimately require surgery. Tendon degeneration is closely associated with irreversible degradation of its collagen structure, leading to changes to its mechanical properties. If these properties could be monitored in vivo, they could be used to quantify the severity of tendonosis and help determine the appropriate treatment. The goal of this cadaveric study was, therefore, to develop and validate ultrasound elasticity imaging (UEI) as a potentially noninvasive technique for quantifying tendon mechanical properties. Five human cadaver feet were mounted in a materials testing system (MTS), while the posterior tibial tendon (PTT) was attached to a force actuator. A portable ultrasound scanner collected 2-D data during loading cycles. Young's modulus was calculated from the strain, loading force, and cross-sectional area of the PTT. Average Young's modulus for the five tendons was (0.45 ± 0.16 GPa) using UEI, which was consistent with simultaneous measurements made by the MTS across the whole tendon (0.52 ± 0.18 GPa). We also calculated the scaling factor (0.12 ± 0.01) between the load on the PTT and the inversion force at the forefoot, a measurable quantity in vivo. This study suggests that UEI could be a reliable in vivo technique for estimating the mechanical properties of the PTT, and as a clinical tool, help guide treatment decisions for advanced PTTD and other tendinopathies.

Original languageEnglish (US)
Article number6985612
Pages (from-to)1179-1184
Number of pages6
JournalIEEE Transactions on Biomedical Engineering
Volume62
Issue number4
DOIs
StatePublished - Apr 1 2015

Fingerprint

Tendons
Elasticity
Ultrasonics
Imaging techniques
Mechanical properties
Materials testing
Elastic moduli
Physical therapy
Collagen
Surgery
Actuators
Degradation

Keywords

  • posterior tibial tendon dysfunction
  • strain imaging
  • Ultrasound elastography

ASJC Scopus subject areas

  • Biomedical Engineering

Cite this

Ultrasound Elasticity Imaging for Determining the Mechanical Properties of Human Posterior Tibial Tendon : A Cadaveric Study. / Gao, Liang; Yuan, Justin S.; Heden, Gregory J.; Szivek, John; Taljanovic, Mihra; Latt, Daniel L; Witte, Russell S.

In: IEEE Transactions on Biomedical Engineering, Vol. 62, No. 4, 6985612, 01.04.2015, p. 1179-1184.

Research output: Contribution to journalArticle

@article{0a65539e71e5455689e6789b02f0c500,
title = "Ultrasound Elasticity Imaging for Determining the Mechanical Properties of Human Posterior Tibial Tendon: A Cadaveric Study",
abstract = "Posterior tibial tendon dysfunction (PTTD) is a common degenerative condition leading to a severe impairment of gait. There is currently no effective method to determine whether a patient with advanced PTTD would benefit from several months of bracing and physical therapy or ultimately require surgery. Tendon degeneration is closely associated with irreversible degradation of its collagen structure, leading to changes to its mechanical properties. If these properties could be monitored in vivo, they could be used to quantify the severity of tendonosis and help determine the appropriate treatment. The goal of this cadaveric study was, therefore, to develop and validate ultrasound elasticity imaging (UEI) as a potentially noninvasive technique for quantifying tendon mechanical properties. Five human cadaver feet were mounted in a materials testing system (MTS), while the posterior tibial tendon (PTT) was attached to a force actuator. A portable ultrasound scanner collected 2-D data during loading cycles. Young's modulus was calculated from the strain, loading force, and cross-sectional area of the PTT. Average Young's modulus for the five tendons was (0.45 ± 0.16 GPa) using UEI, which was consistent with simultaneous measurements made by the MTS across the whole tendon (0.52 ± 0.18 GPa). We also calculated the scaling factor (0.12 ± 0.01) between the load on the PTT and the inversion force at the forefoot, a measurable quantity in vivo. This study suggests that UEI could be a reliable in vivo technique for estimating the mechanical properties of the PTT, and as a clinical tool, help guide treatment decisions for advanced PTTD and other tendinopathies.",
keywords = "posterior tibial tendon dysfunction, strain imaging, Ultrasound elastography",
author = "Liang Gao and Yuan, {Justin S.} and Heden, {Gregory J.} and John Szivek and Mihra Taljanovic and Latt, {Daniel L} and Witte, {Russell S}",
year = "2015",
month = "4",
day = "1",
doi = "10.1109/TBME.2014.2381002",
language = "English (US)",
volume = "62",
pages = "1179--1184",
journal = "IEEE Transactions on Biomedical Engineering",
issn = "0018-9294",
publisher = "IEEE Computer Society",
number = "4",

}

TY - JOUR

T1 - Ultrasound Elasticity Imaging for Determining the Mechanical Properties of Human Posterior Tibial Tendon

T2 - A Cadaveric Study

AU - Gao, Liang

AU - Yuan, Justin S.

AU - Heden, Gregory J.

AU - Szivek, John

AU - Taljanovic, Mihra

AU - Latt, Daniel L

AU - Witte, Russell S

PY - 2015/4/1

Y1 - 2015/4/1

N2 - Posterior tibial tendon dysfunction (PTTD) is a common degenerative condition leading to a severe impairment of gait. There is currently no effective method to determine whether a patient with advanced PTTD would benefit from several months of bracing and physical therapy or ultimately require surgery. Tendon degeneration is closely associated with irreversible degradation of its collagen structure, leading to changes to its mechanical properties. If these properties could be monitored in vivo, they could be used to quantify the severity of tendonosis and help determine the appropriate treatment. The goal of this cadaveric study was, therefore, to develop and validate ultrasound elasticity imaging (UEI) as a potentially noninvasive technique for quantifying tendon mechanical properties. Five human cadaver feet were mounted in a materials testing system (MTS), while the posterior tibial tendon (PTT) was attached to a force actuator. A portable ultrasound scanner collected 2-D data during loading cycles. Young's modulus was calculated from the strain, loading force, and cross-sectional area of the PTT. Average Young's modulus for the five tendons was (0.45 ± 0.16 GPa) using UEI, which was consistent with simultaneous measurements made by the MTS across the whole tendon (0.52 ± 0.18 GPa). We also calculated the scaling factor (0.12 ± 0.01) between the load on the PTT and the inversion force at the forefoot, a measurable quantity in vivo. This study suggests that UEI could be a reliable in vivo technique for estimating the mechanical properties of the PTT, and as a clinical tool, help guide treatment decisions for advanced PTTD and other tendinopathies.

AB - Posterior tibial tendon dysfunction (PTTD) is a common degenerative condition leading to a severe impairment of gait. There is currently no effective method to determine whether a patient with advanced PTTD would benefit from several months of bracing and physical therapy or ultimately require surgery. Tendon degeneration is closely associated with irreversible degradation of its collagen structure, leading to changes to its mechanical properties. If these properties could be monitored in vivo, they could be used to quantify the severity of tendonosis and help determine the appropriate treatment. The goal of this cadaveric study was, therefore, to develop and validate ultrasound elasticity imaging (UEI) as a potentially noninvasive technique for quantifying tendon mechanical properties. Five human cadaver feet were mounted in a materials testing system (MTS), while the posterior tibial tendon (PTT) was attached to a force actuator. A portable ultrasound scanner collected 2-D data during loading cycles. Young's modulus was calculated from the strain, loading force, and cross-sectional area of the PTT. Average Young's modulus for the five tendons was (0.45 ± 0.16 GPa) using UEI, which was consistent with simultaneous measurements made by the MTS across the whole tendon (0.52 ± 0.18 GPa). We also calculated the scaling factor (0.12 ± 0.01) between the load on the PTT and the inversion force at the forefoot, a measurable quantity in vivo. This study suggests that UEI could be a reliable in vivo technique for estimating the mechanical properties of the PTT, and as a clinical tool, help guide treatment decisions for advanced PTTD and other tendinopathies.

KW - posterior tibial tendon dysfunction

KW - strain imaging

KW - Ultrasound elastography

UR - http://www.scopus.com/inward/record.url?scp=84925840366&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=84925840366&partnerID=8YFLogxK

U2 - 10.1109/TBME.2014.2381002

DO - 10.1109/TBME.2014.2381002

M3 - Article

C2 - 25532163

AN - SCOPUS:84925840366

VL - 62

SP - 1179

EP - 1184

JO - IEEE Transactions on Biomedical Engineering

JF - IEEE Transactions on Biomedical Engineering

SN - 0018-9294

IS - 4

M1 - 6985612

ER -