Validation of the VO2000 calorimeter for measuring resting metabolic rate

Vivian Wahrlich, Luiz A. Anjos, Scott B Going, Timothy G Lohman

Research output: Contribution to journalArticle

35 Citations (Scopus)

Abstract

Background & aims: Metabolic carts used in laboratory settings for the measurement of resting metabolism are cumbersome limiting their use in the field. The validity of a newly developed portable calorimeter (Medical Graphics VO2000) under resting conditions was assessed in comparison to a well-established reference system, the DELTATRAC™. Methods: Gas exchange and energy expenditure were measured for 25 min consecutively using the two devices. Values of the last 20 min were averaged and used in the analysis. The order of device for the first subject was randomly chosen and the calorimeters were alternated thereafter. Results: Among 33 subjects, acceptable measures of resting metabolism were obtained in 25 (11 men) aged 20-78 years because eight subjects (three men) either hyperventilated or did not adapt well enough to the facemask. over(V, ̇) O2, over(V, ̇) CO2, and RQ were not significantly different between devices. Small (2.8%) non-clinically relevant mean differences (-0.145±0.341 MJ day-1) were found. Results of the two devices were highly correlated (r=0.95) yielding a more accurate estimate than predictive equations. Conclusions: The VO2000 calorimeter is a valid system to measure resting metabolism but the facemask may not be suitable for some people.

Original languageEnglish (US)
Pages (from-to)687-692
Number of pages6
JournalClinical Nutrition
Volume25
Issue number4
DOIs
StatePublished - Aug 2006

Fingerprint

Basal Metabolism
Equipment and Supplies
Energy Metabolism
Gases

Keywords

  • Calorimeter
  • Deltatrac
  • Indirect calorimetry
  • Oxygen uptake
  • Resting metabolic rate

ASJC Scopus subject areas

  • Critical Care and Intensive Care Medicine
  • Endocrinology, Diabetes and Metabolism
  • Gastroenterology
  • Health Professions(all)
  • Medicine (miscellaneous)

Cite this

Validation of the VO2000 calorimeter for measuring resting metabolic rate. / Wahrlich, Vivian; Anjos, Luiz A.; Going, Scott B; Lohman, Timothy G.

In: Clinical Nutrition, Vol. 25, No. 4, 08.2006, p. 687-692.

Research output: Contribution to journalArticle

@article{b229885ec5104434a6bb97bfd656f5cb,
title = "Validation of the VO2000 calorimeter for measuring resting metabolic rate",
abstract = "Background & aims: Metabolic carts used in laboratory settings for the measurement of resting metabolism are cumbersome limiting their use in the field. The validity of a newly developed portable calorimeter (Medical Graphics VO2000) under resting conditions was assessed in comparison to a well-established reference system, the DELTATRAC™. Methods: Gas exchange and energy expenditure were measured for 25 min consecutively using the two devices. Values of the last 20 min were averaged and used in the analysis. The order of device for the first subject was randomly chosen and the calorimeters were alternated thereafter. Results: Among 33 subjects, acceptable measures of resting metabolism were obtained in 25 (11 men) aged 20-78 years because eight subjects (three men) either hyperventilated or did not adapt well enough to the facemask. over(V, ̇) O2, over(V, ̇) CO2, and RQ were not significantly different between devices. Small (2.8{\%}) non-clinically relevant mean differences (-0.145±0.341 MJ day-1) were found. Results of the two devices were highly correlated (r=0.95) yielding a more accurate estimate than predictive equations. Conclusions: The VO2000 calorimeter is a valid system to measure resting metabolism but the facemask may not be suitable for some people.",
keywords = "Calorimeter, Deltatrac, Indirect calorimetry, Oxygen uptake, Resting metabolic rate",
author = "Vivian Wahrlich and Anjos, {Luiz A.} and Going, {Scott B} and Lohman, {Timothy G}",
year = "2006",
month = "8",
doi = "10.1016/j.clnu.2006.01.002",
language = "English (US)",
volume = "25",
pages = "687--692",
journal = "Clinical Nutrition",
issn = "0261-5614",
publisher = "Churchill Livingstone",
number = "4",

}

TY - JOUR

T1 - Validation of the VO2000 calorimeter for measuring resting metabolic rate

AU - Wahrlich, Vivian

AU - Anjos, Luiz A.

AU - Going, Scott B

AU - Lohman, Timothy G

PY - 2006/8

Y1 - 2006/8

N2 - Background & aims: Metabolic carts used in laboratory settings for the measurement of resting metabolism are cumbersome limiting their use in the field. The validity of a newly developed portable calorimeter (Medical Graphics VO2000) under resting conditions was assessed in comparison to a well-established reference system, the DELTATRAC™. Methods: Gas exchange and energy expenditure were measured for 25 min consecutively using the two devices. Values of the last 20 min were averaged and used in the analysis. The order of device for the first subject was randomly chosen and the calorimeters were alternated thereafter. Results: Among 33 subjects, acceptable measures of resting metabolism were obtained in 25 (11 men) aged 20-78 years because eight subjects (three men) either hyperventilated or did not adapt well enough to the facemask. over(V, ̇) O2, over(V, ̇) CO2, and RQ were not significantly different between devices. Small (2.8%) non-clinically relevant mean differences (-0.145±0.341 MJ day-1) were found. Results of the two devices were highly correlated (r=0.95) yielding a more accurate estimate than predictive equations. Conclusions: The VO2000 calorimeter is a valid system to measure resting metabolism but the facemask may not be suitable for some people.

AB - Background & aims: Metabolic carts used in laboratory settings for the measurement of resting metabolism are cumbersome limiting their use in the field. The validity of a newly developed portable calorimeter (Medical Graphics VO2000) under resting conditions was assessed in comparison to a well-established reference system, the DELTATRAC™. Methods: Gas exchange and energy expenditure were measured for 25 min consecutively using the two devices. Values of the last 20 min were averaged and used in the analysis. The order of device for the first subject was randomly chosen and the calorimeters were alternated thereafter. Results: Among 33 subjects, acceptable measures of resting metabolism were obtained in 25 (11 men) aged 20-78 years because eight subjects (three men) either hyperventilated or did not adapt well enough to the facemask. over(V, ̇) O2, over(V, ̇) CO2, and RQ were not significantly different between devices. Small (2.8%) non-clinically relevant mean differences (-0.145±0.341 MJ day-1) were found. Results of the two devices were highly correlated (r=0.95) yielding a more accurate estimate than predictive equations. Conclusions: The VO2000 calorimeter is a valid system to measure resting metabolism but the facemask may not be suitable for some people.

KW - Calorimeter

KW - Deltatrac

KW - Indirect calorimetry

KW - Oxygen uptake

KW - Resting metabolic rate

UR - http://www.scopus.com/inward/record.url?scp=33746325829&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=33746325829&partnerID=8YFLogxK

U2 - 10.1016/j.clnu.2006.01.002

DO - 10.1016/j.clnu.2006.01.002

M3 - Article

C2 - 16698140

AN - SCOPUS:33746325829

VL - 25

SP - 687

EP - 692

JO - Clinical Nutrition

JF - Clinical Nutrition

SN - 0261-5614

IS - 4

ER -